Answers are grounded in this paper and cite their sources.
A volume algorithm is established which uses the data generated by computed tomography (CT) or ultrasound scans. The algorithm is based on a polyhedral reconstruction of an object of interest. Phantom studies indicate that the algorithm calculated volume is accurate within 3 percent. In a controlled clinical experiment on a laboratory animal, the algorithm calculated volume showed a maximum error of about 5 percent. In less controllable applications to human clinical data, volume calculations showed variations ranging up to 16 percent.
No takes yet. Share an insight, caveat, or question.
Cook et al. (1980) studied this question.