Key result
Transcatheter Nd:YAG laser photocoagulation of the beating canine left ventricle using a balloon-tipped cardioscope is feasible, producing well-demarcated lesions 2.5-9.5 mm deep in 7 of 11 dogs.
Why the study?
Is transcatheter laser ablation of the left ventricle using a balloon-tipped cardioscope feasible in a canine model?
Is transcatheter laser ablation of the left ventricle using a balloon-tipped cardioscope feasible in a canine model?
Transballoon laser photocoagulation of the beating left ventricle is feasible in a canine model, representing a promising preliminary step toward alternative catheter ablative therapies for ventricular tachycardia.
Feasible in canine model; leaves open translation to catheter ablation for human ventricular tachycardia.
The feasibility of transcatheter laser ablation of the canine left ventricle (LV) was tested using a newly developed cardioscope. In 17 anesthetized dogs, a combined laser-endoscope catheter, consisting of an endoscope encased in a 7-French flexible catheter with an inflatable and transparent balloon at the distal end, was introduced into the LV via the carotid artery. A 1064-nm neodymium-yttrium-aluminum-garnet (Nd:YAG) laser was delivered by laser optic fiber, which was introduced through the transport channel and positioned inside the saline-filled balloon. In 16 of 17 dogs, the endocardial surface of the LV was clearly observed. Laser energy totaling 500-5,000 J was applied sequentially in 13 dogs and laser irradiation was completed in all but 2 of the dogs. The excised hearts revealed well-demarcated oval-shaped lesions 2.5-9.5 mm deep in 7 of 11 dogs. Histologic sections revealed coagulation necrosis surrounded by a rim of contraction band necrosis. Thus, transballoon laser photocoagulation of the beating LV is feasible. The newly combined laser-endoscope catheter, which is still in its preliminary stages and needs to be improved to increase the success rate of photocoagulation, appears to be a promising alternative modality for catheter ablative therapy for ventricular tachycardia.
No takes yet. Share an insight, caveat, or question.
Hirao et al. (1997) studied Canine model (n=17). Transcatheter Nd:YAG laser coagulation using a balloon-tipped cardioscope was evaluated on Feasibility of transcatheter laser ablation. Transcatheter Nd:YAG laser photocoagulation of the beating canine left ventricle using a balloon-tipped cardioscope is feasible, producing well-demarcated lesions 2.5-9.5 mm deep in 7 of 11 dogs.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: