Key result
MRI scans in patients with implantable cardiac devices carry a manageable but nonnegligible hazard, with registry data indicating more events with ICDs than pacemakers.
Why the study?
Does MRI cause hazards in patients with implantable cardiac electronic devices?
Does MRI cause hazards in patients with implantable cardiac electronic devices?
MRI scans in patients with cardiac devices have a manageable but nonnegligible hazard, highlighting the need for adequate precautions and further large-scale registry data.
Requires strict MRI safety protocols in device patients; leaves open need for larger registries quantifying ICD-specific risks.
PURPOSE OF REVIEW: To assess the current knowledge about the potential hazard from MRI in patients with devices such as pacemakers and implantable cardioverter defibrillators (ICDs). RECENT FINDINGS: Most data concern 'MRI unsafe' devices, with only a few studies on 'MRI conditional' devices. No 'MRI safe' cardiac devices are currently available. Studies on 'MRI unsafe' devices tend to be small scale and reflect the experience of individual centres; few provide long-term follow-up data. Many newer devices are approved as 'MRI conditional' based on technical simulations or postmarket surveillance studies. With adequate measures taken before performing an MRI scan, reported complication rates are generally low, but there is a nonnegligible residual risk for power-on reset and lead heating. The presence of abandoned, older leads may affect the propensity for lead heating during MRI with newer devices, including those designated 'MRI conditional'. Very little research has been carried out on the hazard from MRI scans in patients with ICDs, but registry data indicate more events with ICDs than with pacemakers. SUMMARY: The limited available data indicate a manageable but not negligible MRI-associated hazard in patients with implantable cardiac devices. Further controlled studies and large, independent registries, particularly in Europe, are needed to provide important safety information.
No takes yet. Share an insight, caveat, or question.
Jung et al. (2014) conducted a review in Patients with implantable cardiac electronic devices (pacemakers and ICDs). MRI scans was evaluated on MRI-associated hazard (complications such as power-on reset and lead heating). MRI scans in patients with implantable cardiac devices carry a manageable but nonnegligible hazard, with registry data indicating more events with ICDs than pacemakers.