Key result
Electromagnetic navigation bronchoscopy magnetic fields caused no interference in 13 patients with ICDs, with 100% of devices correctly detecting and treating induced ventricular fibrillation.
Why the study?
Does electromagnetic navigation bronchoscopy (ENB) cause electromagnetic interference that prevents implanted cardioverter-defibrillators (ICDs) from detecting and treating ventricular fibrillation?
Observational (n=13)
Does electromagnetic navigation bronchoscopy (ENB) cause electromagnetic interference that prevents implanted cardioverter-defibrillators (ICDs) from detecting and treating ventricular fibrillation?
Answers are grounded in this paper and cite their sources.
Electromagnetic navigation bronchoscopy does not interfere with ICD function, including the detection and treatment of ventricular fibrillation, suggesting it is safe for patients with these devices.
AIMS: Electromagnetic navigation bronchoscopy (ENB) (Superdimension) is a diagnostic and therapeutic tool in patients with lung lesions. Very small data are available about potential interference of ENB magnetic field to implanted cardioverter-defibrillators (ICDs) and any documentation of ICD behaviour if a ventricular tachyarrhythmia occurs during ENB is lacking. We tested a number of selected ICDs to assess if any interference occurs by ENB magnetic field on detection of clinical ventricular fibrillation and shock delivery. METHODS AND RESULTS: Thirteen patients undergoing an ICD implantation or elective replacement with a clinical indication to assess the efficacy of defibrillation underwent: (i) real-time telemetric recording from ICD during ENB activation to detect possible noise; (ii) defibrillation test during exposure to ENB board-generated magnetic field. All tested ICDs showed no noise detection at maximum sensitivity level. Induced ventricular fibrillation was correctly detected and cured by implanted device. No change in programmed ICD parameters was induced by exposure to ENB magnetic field. CONCLUSION: All tested ICDs correctly operated and rescued the patients from induced ventricular fibrillation during ENB. Electromagnetic navigation bronchoscopy appears to be safe in heart patients with an ICD; however, close cardiac monitoring of these patients during ENB must be ensured as correct behaviour of all existing ICDs can only be presumed from compliance of the manufacturer to International Standards which establish procedures for electromagnetic interference checking on implantable devices on different ranges of frequency.
No takes yet. Share an insight, caveat, or question.
Magnani et al. (2014) conducted an observational in Patients undergoing ICD implantation or elective replacement (n=13). Electromagnetic navigation bronchoscopy (ENB) magnetic field exposure was evaluated on Interference by ENB magnetic field on detection of clinical ventricular fibrillation and shock delivery. Electromagnetic navigation bronchoscopy magnetic fields caused no interference in 13 patients with ICDs, with 100% of devices correctly detecting and treating induced ventricular fibrillation.