Key result
The Vitalsens VS100 ambulatory monitoring device showed good correlation with staff-recorded heart rates and identified clinical events like paroxysmal AF missed by routine observations.
Why the study?
Does the Vitalsens VS100 ambulatory monitoring device improve the detection of clinical events compared to routine observations in general medical ward patients?
Observational
Does the Vitalsens VS100 ambulatory monitoring device improve the detection of clinical events compared to routine observations in general medical ward patients?
Continuous ambulatory monitoring with the Vitalsens VS100 system in general wards is feasible and detects clinically relevant events missed by intermittent routine observations.
May enhance event detection on general wards; leaves open whether routine adoption improves outcomes.
An evaluation of a newly CE approved bedside monitoring device used in a general hospital ward is presented. This evaluation has shown that it is feasible to use the system within this environment to provide medical staff with supplementary information on patient health, at more frequent intervals than traditional monitoring methods. The physiological data recorded by the body worn device is wirelessly transmitted to a patient management system for storage and display. Good correlation between heart rate values recorded by hospital staff and those recorded by the automated Vitalsens VS100 system was observed. The system has highlighted clinical information that routine observations alone did not readily identify. This can provide clinicians with a better view of the overall health status of the patient. Such medical issues include those witnessed in this study, namely paroxysmal AF, ectopic beats, increasing heart rates recorded prior to a hypoglycaemic event, general high and low heart rate trends and various instances where clinically relevant ECG data has been captured.
No takes yet. Share an insight, caveat, or question.
Harper et al. (2010) conducted an observational in General medical ward patients. Vitalsens VS100 system vs. Routine observations was evaluated on Feasibility and correlation of heart rate values. The Vitalsens VS100 ambulatory monitoring device showed good correlation with staff-recorded heart rates and identified clinical events like paroxysmal AF missed by routine observations.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: