Morphology discrimination in implantable cardioverter-defibrillators: consistency of template match percentage during atrial tachyarrhythmias at different heart rates

View Full Paper
DTDominic A.M.J. TheunsMRMáximo Rivero‐AyerzaDGDick Goedhart

Key Result

Morphology discrimination in ICDs maintained a consistently high template match percentage during SVTs, independent of ventricular cycle length (85.4% for fast vs 87.1% for slow SVTs).

Study Design

Type

Observational (n=88)

Structured PICO

Does heart rate affect the template match percentage of morphology discrimination during supraventricular tachyarrhythmias in patients with dual-chamber ICDs?

P
Population
88 patients with a dual-chamber ICD (St Jude Medical) with 868 stored tachyarrhythmia electrograms
I
Intervention
Morphology discrimination (MD) algorithm evaluation
C
Comparator
Fast (ventricular cycle length 300-350 ms) vs. slow (ventricular cycle length >400 ms) supraventricular tachyarrhythmias
O
Outcome
Template match percentage during SVT at different heart ratessurrogate

Morphology discrimination in ICDs maintains a consistently high template match percentage during SVTs regardless of heart rate, resulting in high specificity for arrhythmia discrimination.

Main Result

Absolute Event Rate: 85.4% vs 87.1%

Abstract

BACKGROUND: Morphology discrimination (MD) in implantable cardioverter-defibrillators (ICDs) is based on the comparison of the ventricular electrogram during tachycardia with a stored reference template obtained during baseline rhythm. However, the effect of heart rate on the template match percentage during supraventricular tachyarrhythmias (SVT) is not known. The purpose of this study was to evaluate the performance of the template match percentage during SVT at different heart rates. METHODS AND RESULTS: Stored electrograms of 868 tachyarrhythmias from 88 patients with a dual-chamber ICD (St Jude Medical, USA) were analysed by the investigators. The effect of heart rate on template match percentage was estimated by regression analysis. For performance measures, data were corrected for multiple episodes in a patient by using the generalized estimating equation method. The mean template match percentage was 86.6 +/- 22.2% (median 100%) for SVT episodes. No significant differences in template match percentage between fast ventricular cycle length (CL) 300-350 ms and slow (ventricular CL >400 ms) SVTs were observed (85.4 +/- 27.0 vs. 87.1 +/- 19.7%). Using nominal settings, MD alone provided sensitivity and specificity of 70.2% and 89.4% overall, respectively. Morphology discrimination in conjunction with rate branch analysis, sudden onset, and stability yielded sensitivity and specificity of 98.5% and 91.2%, respectively. CONCLUSION: Morphology discrimination has a consistently high template match percentage during SVTs, which is independent of ventricular CL. The consistent high match percentage results in high specificity for arrhythmia discrimination.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Theuns et al. (2008) conducted an observational in Patients with a dual-chamber ICD experiencing tachyarrhythmias (n=88). Morphology discrimination vs. Different heart rates (fast vs slow SVT) was evaluated on Template match percentage during SVT. Morphology discrimination in ICDs maintained a consistently high template match percentage during SVTs, independent of ventricular cycle length (85.4% for fast vs 87.1% for slow SVTs).

synapsesocial.com/papers/6a131196020a8e1c07c07984https://doi.org/10.1093/europace/eun194