Key result
Radiofrequency-powered transseptal puncture results in life-threatening descending aorta perforation during LAAC.
Why the study?
Aortic root injury during transseptal puncture is a recognized complication, but injury to the descending aorta had not previously been described.
Case Report (n=1)
Radiofrequency-powered systems for transseptal puncture can cause life-threatening descending aorta perforation, highlighting the need for caution, particularly in patients with an interatrial septum aneurysm.
May warrant procedural caution in LAA closure; leaves open optimal imaging and technique refinements.
BACKGROUND: Aortic root injury during transseptal puncture (TSP) is a rare but well-recognized complication. However, injury to the descending aorta has not previously been described in this context. CASE SUMMARY: A 57-year-old woman underwent left atrial appendage closure. During TSP using the VersaCross Connect radiofrequency-powered system and a 15-F steerable sheath, an unrecognized perforation of the descending aorta occurred. Despite repeated transthoracic echocardiograms showing no pericardial effusion and initial hemodynamic stability, she subsequently developed a mediastinal hematoma and bilateral pleural effusion. Management required an emergency percutaneous thoracic endovascular aortic repair, followed by surgical evacuation of the hematoma. DISCUSSION: Excessive tenting during radiofrequency-powered TSP, related to interatrial septum aneurysm, resulted in injury to the descending aorta with life-threatening consequences. TAKE-HOME MESSAGE: Radiofrequency-powered systems for TSP have enhanced tissue penetration capability and should therefore be used with caution, particularly in the presence of an interatrial septum aneurysm and atypical echocardiographic windows.
No takes yet. Share an insight, caveat, or question.
Bini et al. (2026) conducted a case report in Left atrial appendage closure (n=1). Transseptal puncture using VersaCross Connect radiofrequency-powered system was evaluated on Descending aorta perforation. Radiofrequency-powered transseptal puncture resulted in an unrecognized, life-threatening perforation of the descending aorta in a 57-year-old woman undergoing left atrial appendage closure.