Key result
Mobile 6-lead ECG accurately measures QTc with a ~12 ms difference versus conventional 12-lead ECG.
Why the study?
Monitoring QTc with repeated 12-lead ECGs is cumbersome, and single-lead mobile ECG devices have variable accuracy, suggesting a multilead mobile ECG device may be more accurate.
Does a 6-lead mobile ECG device accurately measure QTc compared to a conventional 12-lead ECG in cardiology outpatients?
Observational (n=234)
Unblinded
Yes
Does a 6-lead mobile ECG device accurately measure QTc compared to a conventional 12-lead ECG in cardiology outpatients?
A 6-lead mobile ECG device provides accurate QTc measurements compared to a 12-lead ECG, particularly in lead II, supporting its potential use for ambulatory QTc monitoring.
Answers are grounded in this paper and cite their sources.
Supports mobile QTc monitoring in outpatients; leaves open need for prospective validation before routine use.
INTRODUCTION: Ambulatory assessment of the heart rate-corrected QT interval (QTc) can be of diagnostic value, for example in patients on QTc-prolonging medication. Repeating sequential 12-lead electrocardiograms (ECGs) to monitor the QTc is cumbersome, but mobile ECG (mECG) devices can potentially solve this problem. As the accuracy of single-lead mECG devices is reportedly variable, a multilead mECG device may be more accurate. METHODS: This prospective dual-centre study included outpatients visiting our cardiology clinics for any indication. Participants underwent an mECG recording using a smartphone-enabled 6‑lead mECG device immediately before or immediately after a conventional 12-lead ECG recording. Multiple QTc values in both recordings were manually measured in leads I and II using the tangent method and subsequently compared. RESULTS: In total, 234 subjects were included (mean ± standard deviation (SD) age: 57 ± 17 years; 58% males), of whom 133 (57%) had cardiac disease. QTc measurement in any lead was impossible due to artefacts in 16 mECGs (7%) and no 12-lead ECGs. Mean (± SD) QTc in lead II on the mECG and 12-lead ECG was 401 ± 30 and 406 ± 31 ms, respectively. Mean (± SD) absolute difference in QTc values between both modalities was 12 ± 9 ms (r = 0.856; p < 0.001). In 55% of the subjects, the absolute difference between QTc values was < 10 ms. CONCLUSION: A 6-lead mECG allows for QTc assessment with good accuracy and can be used safely in ambulatory QTc monitoring. This may improve patient satisfaction and reduce healthcare costs.
No takes yet. Share an insight, caveat, or question.
Bergeman et al. (2022) conducted an observational in Cardiology outpatients (n=234). 6-lead mobile ECG (KardiaMobile 6L) vs. Conventional 12-lead ECG was evaluated on Mean QTc interval in lead II (ms) (r = 0.856, p=<0.001). A 6-lead mobile ECG device accurately measured the QTc interval compared to a conventional 12-lead ECG, demonstrating a strong correlation (r = 0.856) and a mean absolute difference of 12 ms in lead II.