Key result
Patient-recorded 6-lead mobile ECG enables accurate rhythm and QTc evaluation in hospitalized patients.
Why the study?
Minimizing patient contact during the COVID-19 era and the need for increased QT/QTc interval monitoring motivated evaluation of contactless mobile ECG technology.
Can a six-lead contactless mobile ECG device provide reliable QT/QTc interval measurements compared to standard 12-lead ECGs in hospitalized patients?
Observational (n=4)
No
Can a six-lead contactless mobile ECG device provide reliable QT/QTc interval measurements compared to standard 12-lead ECGs in hospitalized patients?
A six-lead mobile ECG device allows hospitalized patients to successfully record their own ECGs for reliable QT/QTc monitoring, minimizing healthcare worker contact.
Patient self-recorded 6L ECGs feasible for QTc monitoring; hypothesis-generating and should not yet change practice.
Mobile electrocardiograms (ECGs) (mECGs) using smartphone applications are an emerging technology. In the coronavirus disease 2019 (COVID-19) era, minimizing patient contact has gained increasing importance. Additionally, increased QT/corrected QT (QTc) monitoring has concurrently been required. The KardiaMobile 6L ECG device, cleared by the United States Food and Drug Administration (FDA) for recording ECGs, along with the KardiaStation tablet application is a platform (AliveCor, Mountain View, CA, USA) that addresses these two issues. A team of residents, fellows, hospitalists, and cardiologists identified inpatients in need of QT/QTc interval monitoring to pilot the adoption of a system composed of a KardiaMobile 6L ECG device with the accompanying KardiaStation tablet application. Concurrent standard ECGs provided validation. Adoption and performance issues were recorded. Four patients agreed to participate in QT/QTc interval monitoring, three of whom were positive for severe acute respiratory syndrome coronavirus 2 viral infection. After basic instructions were given to the patients and their clinical nurses, all patients recorded mECGs successfully. Patients were able to record their own mECG tracings at least once without any assistance. The 12-lead ECGs and mECGs each showed the correct rhythm, and the measured QTc intervals on each modality were consistently acceptable (< 500 ms). Contactless ECGs were successfully uploaded to KardiaStation for QT/QTc interval measurement and archiving. In this study, we showed that an FDA-cleared product, KardiaMobile 6L, has the ability to provide high-quality contactless ECGs for reliable QT/QTc interval measurements. Hospitalized patients were able to perform recordings when requested after receiving simple instructions at the time of first use. This technology has applications during the COVID-19 pandemic and beyond.
No takes yet. Share an insight, caveat, or question.
Frisch et al. (2021) conducted an observational in Inpatients requiring intermittent ECG monitoring (including COVID-19) (n=4). KardiaMobile 6L mobile electrocardiogram vs. Standard 12-lead ECG was evaluated on Feasibility of obtaining contactless recordings and qualitative comparison with standard 12-lead ECG. Hospitalized patients successfully recorded their own 6-lead mobile ECGs using the KardiaMobile 6L device after simple instructions, providing accurate rhythm and QTc interval measurements.