Why the study?
The study was conducted to compare early clinical outcomes after TAVI across three consecutive generations of self-expanding valves.
Does transcatheter aortic valve implantation with newer generation self-expanding valves (Evolut R or Evolut PRO) reduce the need for permanent pacemaker implantation and paravalvular leakage compared to the first-generation CoreValve in patients with severe aortic stenosis?
Observational (n=368)
No
Does transcatheter aortic valve implantation with newer generation self-expanding valves (Evolut R or Evolut PRO) reduce the need for permanent pacemaker implantation and paravalvular leakage compared to the first-generation CoreValve in patients with severe aortic stenosis?
OBJECTIVE: To compare early clinical outcomes after transcatheter aortic valve implantation (TAVI) with three consecutive generations of self-expanding valves (SEVs). METHODS: Clinical endpoints of consecutive patients who underwent TAVI with CoreValve, Evolut R or Evolut PRO were included in a prospective database. RESULTS: TAVI was performed with CoreValve (n = 116), Evolut R (n = 160) or Evolut PRO (n = 92). Evolut R and Evolut PRO showed a tendency towards lower permanent pacemaker implantation (PPI) rates compared to CoreValve (CoreValve 27% vs Evolut R 16% vs Evolut PRO 18%, p = 0.091). By multivariable regression analysis CoreValve had a significantly higher risk for PPI (odds ratio (OR) 2.79, 95% confidence interval (CI) 1.31-5.94, p = 0.008) compared to Evolut R, while Evolut R and PRO were similar. Severe paravalvular leakage (PVL) occurred only with CoreValve, but no significant difference was observed in moderate PVL (10% vs 8% vs 6%, p = 0.49). CoreValve had a tendency towards a higher risk for more-than-mild PVL as compared with the Evolut platform (R + PRO) (OR 2.46, 95% CI 0.98-6.16, p = 0.055). No significant differences in all-cause mortality (7% vs 4% vs 1%, p = 0.10), stroke (6% vs 3% vs 2%, p = 0.21) or major vascular complications (10% vs 12% vs 4%, p = 0.14) were observed. CONCLUSIONS: TAVI with self-expanding valves was safe, and device iterations may result in a lower need for PPI. More-than-mild PVL seemed to occur less often with repositionable technology.
No takes yet. Share an insight, caveat, or question.
Kroon et al. (2021) conducted an observational in Severe aortic stenosis (n=368). Evolut R vs. CoreValve was evaluated on Permanent pacemaker implantation at 30 days (OR 2.79, 95% CI 1.31-5.94, p=0.008).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: