Review and measurement characterize electrical conductivity of various tissues and fluids, suggesting methods for improvement.
A two-pronged approach, review and measurement, has been adopted to characterize the conductivity of tissues at frequencies below 1 MHz. The review covers data published in the last decade and earlier data not included in recent reviews. The measurements were carried out on pig tissue, in vivo, and pig body fluids in vitro. Conductivity data have been obtained for skeletal and myocardial muscle, liver, skull, fat, lung and body fluids (blood, bile, CSF and urine). A critical analysis of the data highlights their usefulness and limitations and enables suggestions to be made for measuring the electrical properties of tissues.
No takes yet. Share an insight, caveat, or question.
Gabriel et al. (2009) studied this question.