Key result
Real-time three-dimensional color flow and spectral Doppler ultrasound imaging was feasible using a 12 French catheter-mounted two-dimensional array transducer in phantoms and an in vivo sheep model.
Why the study?
Is real-time three-dimensional Doppler intracardiac echocardiography feasible in phantoms and an open chest sheep model?
Is real-time three-dimensional Doppler intracardiac echocardiography feasible in phantoms and an open chest sheep model?
Real-time 3D Doppler intracardiac echocardiography is feasible using a catheter-mounted 2D array transducer in preclinical models.
Preclinical feasibility supports device iteration; leaves open human translation of 3D Doppler ICE.
Using catheter-mounted two-dimensional array transducers, we have obtained real-time three-dimensional color flow and spectral Doppler ultrasound images in phantoms. We have constructed several transducers operating at 5 MHz with up to 137 channels inside a 12 French catheter lumen. The transducer configuration may be side scanning or beveled with respect to the long axis of the catheter lumen. We have also included six electrodes to acquire simultaneous electrocardiograms. We have measured Doppler signals in two phantoms, a string phantom and a pulsatile flow phantom. Using an open chest sheep model, we inserted the catheter into the cardiac chambers to study the utility of in vivo 3D intracardiac Doppler echocardiography.
No takes yet. Share an insight, caveat, or question.
Light et al. (2002) studied this question. Real-time three-dimensional Doppler intracardiac echocardiography was evaluated. Real-time three-dimensional color flow and spectral Doppler ultrasound imaging was feasible using a 12 French catheter-mounted two-dimensional array transducer in phantoms and an in vivo sheep model.