Key result
Real-time three dimensional intracardiac echocardiography successfully produced up to 60 3D scans per second and provided real-time 3D images of anatomical landmarks in an open chest sheep model.
Why the study?
Is real-time three dimensional intracardiac echocardiography feasible for imaging anatomical landmarks in an open chest sheep model?
Is real-time three dimensional intracardiac echocardiography feasible for imaging anatomical landmarks in an open chest sheep model?
Real-time 3D intracardiac echocardiography is feasible for visualizing key anatomical landmarks, suggesting potential utility for guiding electrophysiological procedures.
Answers are grounded in this paper and cite their sources.
Supports preclinical validation of real-time 3D ICE; leaves open human feasibility for procedural guidance.
We have tested the feasibility of real-time three dimensional intracardiac echocardiography for guidance of interventional electrophysiology studies. We have constructed 2D array ultrasound transducers operating at 5MHz with 64 channels in a 12 F catheter, and of 137 channels in an 8 F catheter with a 12 F flexible interconnect. The system produces up to 60 3D scans per second and features real-time volume rendering. Using an open chest sheep model, we obtained real-time 3D images of anatomical landmarks, including the pulmonary veins and coronary sinus, which are of value in electrophysiological procedures.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2002) studied Interventional electrophysiology studies. Real-time three dimensional intracardiac echocardiography was evaluated on Feasibility of obtaining real-time 3D images of anatomical landmarks. Real-time three dimensional intracardiac echocardiography successfully produced up to 60 3D scans per second and provided real-time 3D images of anatomical landmarks in an open chest sheep model.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: