Key result
Intramyocardial fat alters electric field distribution and PFA lesion size due to lower electrical conductivity.
Why the study?
There was no formal physical-based analysis describing how fat can affect the electric field distribution during cardiac pulsed field ablation.
Does intramyocardial fat alter the electric field distribution during Pulsed Field Ablation in computer models?
Does intramyocardial fat alter the electric field distribution during Pulsed Field Ablation in computer models?
Computer modeling demonstrates that intramyocardial fat significantly alters the electric field distribution and lesion shape during Pulsed Field Ablation due to its low electrical conductivity.
Answers are grounded in this paper and cite their sources.
May alter PFA lesion prediction in fatty hearts; leaves open need for in vivo validation and protocol adjustments.
Even though the preliminary experimental data suggests that cardiac Pulsed Field Ablation (PFA) could be superior to radiofrequency ablation (RFA) in terms of being able to ablate the viable myocardium separated from the catheter by collagen and fat, as yet there is no formal physical-based analysis that describes the process by which fat can affect the electric field distribution. Our objective was thus to determine the electrical impact of intramyocardial fat during PFA by means of computer modeling. Computer models were built considering a PFA 3.5-mm blunt-tip catheter in contact with a 7-mm ventricular wall (with and without a scar) and a 2-mm epicardial fat layer. High voltage was set to obtain delivered currents of 19, 22 and 25 A. An electric field value of 1000 V/cm was considered as the lethal threshold. We found that the presence of fibrotic tissue in the scar seems to have a similar impact on the electric field distribution and lesion size to that of healthy myocardium only. However, intramyocardial fat considerably alters the electrical field distribution and the resulting lesion shape. The electric field tends to peak in zones with fat, even away from the ablation electrode, so that 'cold points' (i.e. low electric fields) appear around the fat at the current entry and exit points, while 'hot points' (high electric fields) occur in the lateral areas of the fat zones. The results show that intramyocardial fat can alter the electric field distribution and lesion size during PFA due to its much lower electrical conductivity than that of myocardium and fibrotic tissue.
No takes yet. Share an insight, caveat, or question.
Pérez et al. (2023) studied this question. Pulsed Field Ablation (PFA) was evaluated on Electric field distribution and lesion size. Intramyocardial fat considerably alters the electric field distribution (lethal threshold 1000 V/cm) and lesion size during Pulsed Field Ablation due to its lower electrical conductivity.
Synapse has enriched 10 related papers. Consider them for comparative context: