Key result
The WKflow pulse contour method accurately measured beat-to-beat stroke volume during intra-aortic balloon pumping compared to the conductance method, with a bias of 1.04 ml and precision of 10.9%.
Why the study?
To evaluate a new Windkessel model-based pulse contour method (WKflow) for calculating stroke volume in patients undergoing intra-aortic balloon pumping.
Does the Windkessel model based pulse contour method accurately measure stroke volume compared to the left ventricular conductance volume method in patients with low ejection fraction undergoing intra-aortic balloon pumping?
Observational (n=12)
No
Does the Windkessel model based pulse contour method accurately measure stroke volume compared to the left ventricular conductance volume method in patients with low ejection fraction undergoing intra-aortic balloon pumping?
p-value: p=0.007
The Windkessel model based pulse contour method is interchangeable with the left ventricular conductance volume method for measuring absolute stroke volume and tracking changes in patients undergoing IABP.
May support pulse contour SV monitoring in IABP; leaves open prospective validation before routine use.
Evaluation of a new Windkessel model based pulse contour method (WKflow) to calculate stroke volume in patients undergoing intra-aortic balloon pumping (IABP). Preload changes were induced by vena cava occlusions (VCO) in twelve patients undergoing cardiac surgery to vary stroke volume (SV), which was measured by left ventricular conductance volume method (SVlv) and WKflow (SVwf). Twelve VCO series were carried out during IABP assist at a 1:2 ratio and seven VCO series were performed with IABP switched off. Additionally, SVwf was evaluated during nine episodes of severe arrhythmia. VCO’s produced marked changes in SV over 10–20 beats. 198 paired data sets of SVlv and SVwf were obtained. Bland–Altman analysis for the difference between SVlv and SVwf during IABP in 1:2 mode showed a bias (accuracy) of 1.04 ± 3.99 ml, precision 10.9% and limits of agreement (LOA) of − 6.94 to 9.02 ml. Without IABP bias was 0.48 ± 4.36 ml, precision 11.6% and LOA of − 8.24 to 9.20 ml. After one thermodilution calibration of SVwf per patient, during IABP the accuracy improved to 0.14 ± 3.07 ml, precision to 8.3% and LOA to − 6.00 to + 6.28 ml. Without IABP the accuracy improved to 0.01 ± 2.71 ml, precision to 7.5% and LOA to − 5.41 to + 5.43 ml. Changes in SVlv and SVwf were directionally concordant in response to VCO’s and during severe arrhythmia. (R2 = 0.868). The SVwf and SVlv methods are interchangeable with respect to measuring absolute stroke volume as well as tracking changes in stroke volume. The precision of the non-calibrated WKflow method is about 10% which improved to 7.5% after one calibration per patient.
No takes yet. Share an insight, caveat, or question.
Jansen et al. (2019) conducted an observational in Low ejection fraction undergoing cardiac surgery requiring intra-aortic balloon pumping (n=12). WKflow pulse contour method vs. Left ventricular conductance volume method was evaluated on Bias between stroke volume by conductance (SVlv) and WKflow (SVwf) during IABP in 1:2 mode (95% CI -6.94 to 9.02, p=0.007). The WKflow pulse contour method accurately measured beat-to-beat stroke volume during intra-aortic balloon pumping compared to the conductance method, with a bias of 1.04 ml and precision of 10.9%.