Why the study?
How does the number of surface electrodes affect the accuracy of cardiac dipole moment estimation in normal children?
How does the number of surface electrodes affect the accuracy of cardiac dipole moment estimation in normal children?
Accurate estimation of the cardiac dipole moment using surface potential integration requires at least 100 surface electrodes to keep relative errors below 15%.
Answers are grounded in this paper and cite their sources.
Fewer than 100 electrodes may yield >15% error in pediatric dipole estimates; leaves open optimal configurations for clinical mapping systems.
This study reports on the errors in the calculation of a heart dipole introduced by deleting individual electrode contributions to a surface potential map. The potential at 150 points on each of 6 normal children was available at several instants of the QRS cycle. Although it was impossible to test all possible sets of deleted points, or all possible methods of compensatory weighting of the residual points, the results indicate that a number of electrodes at least as great as 100 is required to produce consistent relative errors of less than 15 percent, if surface potential integration is employed in determining the heart's total dipole moment.
No takes yet. Share an insight, caveat, or question.
Weiss et al. (1970) studied this question.