Key result
A cooled intraesophageal balloon effectively protects the esophageal lumen during RF ablation, but temperatures reached 2 mm away may still risk injury to inner esophageal layers.
Why the study?
Does a cooled intraesophageal balloon prevent thermal injury during RF cardiac ablation in an agar phantom model?
Does a cooled intraesophageal balloon prevent thermal injury during RF cardiac ablation in an agar phantom model?
While a cooled intraesophageal balloon protects the esophageal lumen during RF ablation, deeper tissue layers may still be at risk of thermal injury.
Answers are grounded in this paper and cite their sources.
Cooled balloons may leave deeper esophageal layers vulnerable during RF ablation; leaves open need for human safety studies.
UNLABELLED: Cooled Balloon Prevents Thermal Injury During RF Ablation. INTRODUCTION: The use of a cooled intraesophageal balloon has recently been proposed to minimize the risk of thermal injury in the esophagus during radiofrequency (RF) ablation of the left atrium. However, the capacity of this device to adequately protect the esophagus under different procedural and anatomical conditions remains unknown. METHODS AND RESULTS: An agar phantom-based model was built that provided temperature readings not only on the cooled balloon (T(b)) but also at a hypothetical point between the esophageal lumen and myocardium at a distance of 2 mm (T(2-mm)). The RF ablations were conducted considering two anatomical factors (total distance between the electrode and balloon and flow rate around the electrode) and two procedural factors (angle and pressure between the electrode and agar surface). The results show that most of the parameters studied have no significant influence on the temperature measured on the cooled balloon (T(b)), the exception being a variation in the flow rate, which was found to influence the temperature. On the other hand, T(2-mm) was affected to a great extent by all the factors considered, the smallest influence being that of the contact pressure. The results also suggest that when an intraesophageal balloon is employed, the applied power is not a good predictor either of the temperature on the balloon or of the temperature measured at a distance 2 mm away. CONCLUSION: The results suggest that a cooled intraesophageal balloon provides effective thermal protection of the esophageal lumen. However, under certain circumstances, the temperature reached at a distance 2 mm away could possibly put at risk the integrity of the inner layers of the esophagus.
No takes yet. Share an insight, caveat, or question.
Lequerica et al. (2008) studied Thermal injury during RF cardiac ablation. Cooled intraesophageal balloon was evaluated on Temperature on the cooled balloon (T(b)) and at a hypothetical point 2 mm away (T(2-mm)). A cooled intraesophageal balloon effectively protects the esophageal lumen during RF ablation, but temperatures reached 2 mm away may still risk injury to inner esophageal layers.