Key result
Concomitant CRT in LVAD recipients is not associated with improved survival despite ~14% device complications.
Why the study?
The effects of cardiac resynchronization therapy on survival and ventricular arrhythmias in patients with left ventricular assist devices remain uncertain.
Does cardiac resynchronization therapy improve survival or reduce ventricular arrhythmias in patients with a left ventricular assist device?
Cohort (n=652)
Yes
Does cardiac resynchronization therapy improve survival or reduce ventricular arrhythmias in patients with a left ventricular assist device?
Answers are grounded in this paper and cite their sources.
Concomitant cardiac resynchronization therapy in LVAD patients does not confer a survival benefit or reduce ventricular arrhythmias, but is associated with notable device-related complications.
A significant number of patients with left ventricular assist devices (LVAD) are also treated with cardiac resynchronization therapy (CRT). However, the effects of CRT on LVAD-recipient survival and ventricular arrhythmias (VAs) remain uncertain. To address this issue, we analyzed patients from the multicenter ASSIST-ICD registry using a 1:1 propensity score-matched cohort to evaluate the association between CRT and all-cause mortality. Secondary outcomes included cardiovascular/noncardiovascular mortality, and early (≤30 days postimplant) or late VAs. Among 652 LVAD recipients, 198 patients (30.4%) had CRT before LVAD implantation, followed for a median duration of 9.1 months (2.5-22.1). Cardiac resynchronization therapy patients were older with higher rates of atrial fibrillation, history of VAs, ischemic cardiomyopathy, and longer heart failure duration. After propensity matching, CRT use was not associated with improved survival (hazard ratio [HR]: 1.31 [0.94-1.82]), cardiovascular mortality (HR: 1.14 [0.68-1.90]), noncardiovascular mortality (HR: 0.88 [0.56-1.40]), or incidence of early (odds ratio [OR]: 0.71 [0.43-1.16]) or late (OR: 1.24 [0.73-2.11]) VAs. However, 27 CRT patients (13.6%) experienced device-related complications. These findings suggest that concomitant CRT with LVAD is not associated with survival benefit, despite notable device-related complications, warranting future randomized trials to clarify the clinical impact of pre-existing CRT in LVAD patients.
No takes yet. Share an insight, caveat, or question.
Cherbi et al. (2025) conducted a cohort in Left ventricular assist device (LVAD) recipients (n=652). Cardiac resynchronization therapy (CRT) vs. No CRT was evaluated on All-cause mortality (HR 1.31, 95% CI 0.94-1.82). Concomitant cardiac resynchronization therapy in LVAD recipients was not associated with improved survival (HR 1.31; 95% CI 0.94-1.82) despite a 13.6% rate of device-related complications.
Synapse has enriched 10 related papers. Consider them for comparative context: