This is the authors' abstract. We don't add key points for this paper.
Reliable acoustic coupling in a non-handheld mode and reducing the form factor of electronics are specific challenges in making ultrasound wearable. Applications relying on a large field of view (such as tracking of large muscles) induce a need for a large element count to achieve high image quality. In our work, we developed a 256-element linear array for imaging of abdominal muscles with four integrated custom-developed 8:32 multiplexer Integrated Circuits (ICs), allowing the array to be driven by our compact 32 ch electronics. The system is optimized for flexible use in R&D applications and allows adjustable transmit voltages (up to +/−100 V), arbitrary delay patterns, and 12-bit analog-to-digital conversion (ADC) with up to 50 MSPS and wireless (21.6 MBit/s) or USB link. Image metrics (SLL, FWHM) were very similar to a fully populated array driven with a 256 ch system. The contrast allowed imaging of lesions down to 7 cm in the phantom. In a first in-vivo study, we demonstrated reliable acoustic contact even during exercise and were able to visualize deep abdominal muscles such as the TrA. In combination with a muscle tracking algorithm, the change of thickness of the TrA during SSE could be monitored, demonstrating the potential of the approach as biofeedback for physiotherapy training.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Speicher et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: