Answers are grounded in this paper and cite their sources.
The sensitivity of an energy-compensated Geiger-Müller counter to neutrons with approximate energies of 0.6, 5 and 15 MeV, relative to its sensitivity to gamma rays (kU-value), has been determined at different centres in Europe. The time-of-flight, angular difference, associated-particle coincidence and lead filtration technique have been compared, using the same Geiger-Müller counter. For 15 MeV neutrons, a consistent kU-value of 1.9% was obtained. For the lower neutron energies, lower kU values were observed with reasonable agreement.
No takes yet. Share an insight, caveat, or question.
Mijnheer et al. (1982) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: