Proposed algorithm measuring pulse transit time from ECG and PPG achieves ~98% sensitivity for PPG-foot detection.
Why the study?
Pulse transit time is a physiological parameter and direct indicator of cardiovascular diseases, but low-complexity methods for its measurement are needed.
A newly proposed low-complexity algorithm for measuring pulse transit time from ECG and PPG signals demonstrated high sensitivity and accuracy (>94%) for peak and foot detection.
Answers are grounded in this paper and cite their sources.
Supports technical validation of ECG-PPG algorithms; leaves open clinical adoption for pulse transit time assessment.
Pulse transit Time (PTT) is a physiological parameter that is based on characteristics of the pulse waveform, a direct indicator of Cardiovascular Diseases (CVD). The (PTT) is defined as the time taken for the arterial pulse to travel from the heart to a peripheral site. It is commonly derived from Electrocardiogram (ECG) and Photoplethysmogram (PPG) signal calculations and is calculated as the interval between the peak of the electrocardiogram (ECG) R-wave and a time point on the photoplethysmogram (PPG). In this study, we propose a new and lowcomplexity algorithm for the Pulse transit time (PTT) measurement, using these two signals and detecting (PTT- foot) and (PTT- peak). We built a 37 subjects dataset containing a simultaneous recording of the (ECG) and (PPG). The calculation of (PTT) consists of detecting the peak and foot points of a (PPG) and the R-peak of the (ECG) signal. Intermediate operations such as normalization and thresholding to detect the local maxima and minima, are processed on noisy signals, this algorithm is improved by a windowing temporal analysis. The obtained results are promising for the first step. The average sensitivity (SEN) and accuracy (ACC) obtained were (97.5%, and 96.82%) respectively for R-peaks detection and respectively (97.77%, and 97.64%) for (PPG-peak) detection. The sensitivity (SEN) and accuracy (ACC) of (PPG- foot) detection were (98.33%, and 94.14%).
No takes yet. Share an insight, caveat, or question.
Lilia et al. (2023) studied this question. New algorithm for measuring Pulse Transit Time (PTT) was evaluated on Sensitivity of PPG-foot detection. The proposed algorithm for measuring pulse transit time from ECG and PPG signals achieved an average sensitivity of 98.33% and accuracy of 94.14% for PPG-foot detection.