Key result
Placing one defibrillation electrode in the persistent left superior vena cava perpendicular to the right ventricular electrode allowed successful transvenous defibrillation without a subcutaneous patch.
Case Report (n=1)
No
Answers are grounded in this paper and cite their sources.
In patients with a persistent left superior vena cava, placing a defibrillation electrode in the anomalous vein perpendicular to the right ventricular electrode can achieve adequate defibrillation thresholds without needing a subcutaneous patch.
The implantation of a transvenous cardioverter defibrillator (PCD 7217B) was performed in a patient with a persistent left superior vena cava. The defibrillation electrodes were positioned in the right ventricle and the superior vena cava via the right subclavian vein. A subcutaneous patch had to be implanted at the left lateral chest wall to achieve sufficient defibrillation thresholds. Three weeks later the system had to be removed because of a generator pocket infection. During the second implantation we placed one electrode in the persistent left superior vena cava perpendicular to the electrode in the right ventricle. Using this configuration transvenous defibrillation was possible without an additional subcutaneous patch.
No takes yet. Share an insight, caveat, or question.
MATTKE et al. (1995) conducted a case report in Persistent left superior vena cava requiring defibrillator implantation (n=1). Transvenous cardioverter defibrillator implantation with electrode in persistent left superior vena cava was evaluated on Successful transvenous defibrillation without an additional subcutaneous patch. Placing one defibrillation electrode in the persistent left superior vena cava perpendicular to the right ventricular electrode allowed successful transvenous defibrillation without a subcutaneous patch.
Synapse has enriched 10 related papers. Consider them for comparative context: