Key result
Microcontroller ECG prototype matches manual MET calculations and accurately assesses ST segments.
Answers are grounded in this paper and cite their sources.
A prototype device and mobile application successfully automated MET calculation from the Rockport Walking Fitness Test and assessed ST segments for abnormality.
Cardiovascular diseases have always been among the top causes of death. Thus, there were variety of research focuses on cardiac stress test and on self-stress test as well. However, the portable device existed for heart monitoring were still insufficient where most device are for fitness monitoring rather than functional capacity. Thus, this paper determines heart condition based on New York Heart Classification (NYHA) for The Rockport Walking Fitness Test (TRWFT). TRWFT gave out parameters to calculate maximum oxygen consumption (VO 2 max) to find metabolic equivalent (MET) value. The MET values were compared with NYHA functional capacity classes. Hence, the comparison simplified the heart condition to the urgency to visit hospitals. Plus, ST segments of MIT-BIH ECG database were extracted and assessed for abnormality. Accordingly, a mobile application was developed. The prototype is built using Genuino 101 microcontroller, an AD8232 ECG module with leads to sense heart pulse, and a SD card module. Meanwhile, simulation test on the prototype shows that the prototype succeeds in produce the same values with manual calculation and managed to assess ST segments as programmed. In conclusion, a prototype to demonstrate the implementation of TRFWT and prediction of cardiac condition based on MET value is successful.
No takes yet. Share an insight, caveat, or question.
Yassin et al. (2023) studied Cardiovascular diseases. Prototype device and mobile application for TRWFT and ST segment assessment vs. Manual calculation was evaluated on Accuracy of MET value calculation and ST segment assessment. A developed prototype using a microcontroller and ECG module successfully produced the same MET values as manual calculation and managed to assess ST segments as programmed.
Synapse has enriched 10 related papers. Consider them for comparative context: