Key result
High-power PVI results in steam pops in ~2% of patients, predicted by significant impedance changes.
Why the study?
Despite the cooling effect of contact-force porous catheters, steam pop risk remains a major concern during high-power CPVI.
What are the prevalence and predictors of steam pops during high-power circumferential pulmonary vein isolation?
Case-Control (n=701)
What are the prevalence and predictors of steam pops during high-power circumferential pulmonary vein isolation?
An impedance drop >12 Ω during high-power pulmonary vein isolation is a significant predictor of steam pops, which occur in approximately 1.57% of cases.
Answers are grounded in this paper and cite their sources.
Impedance drop >12 Ω may flag steam pop risk during high-power PVI; leaves open prospective validation before guiding ablation technique.
Despite the good cooling effect of the contact-force porous catheter, the risk of steam pops (SP) remains one of the major concerns in high-power circumferential pulmonary vein isolation (CPVI). This study aimed to investigate the prevalence, predictors and possible mechanisms of SPs in CPVI. Patients experiencing SPs in de novo high-power CPVI were 1:3 matched by non-SP patients with gender, age (±5 years) and left atrial diameter (LAD) (±5 mm) to compare the ablation parameters of SP and non-SP lesions. Catheter tip displacement (Tipdisp) was compared between “edge-of-ridge” and “PV-side-of-ridge” placement at anterior and roof segments of the left pulmonary vein (PV). SPs occurred in 11 (1.57%) of 701 patients, including 6 at the antero-superior left PV, 2 at the roof, 1 at the postero-superior left PV, 1 at the bottom left PV and 1 at the antero-superior aspect of the right PV. There was significantly shorter RF delivery duration (13.9 ± 6.3 vs. 23.3 ± 6.0 s), greater Δimpedance (17.6 ± 6.7 vs. 6.7 ± 4.1 Ω) and lower ablation index (357.7 ± 68.8 vs. 430.2 ± 30.7) in SP patients than those in non-SP patients. Δimpedance >12 Ω during ablation could predict SP occurrence. Tipdisp was greater in “PV-side-of-ridge” than that in “edge-of -ridge” placement (3.2 ± 1.6 mm vs. 2.0 ± 0.8 mm) at antero-superior and roof segments of the left PV. The prevalence of SP was 1.57% in high-power CPVI procedures, with the most common site at the antero-superior segment of the left PV. Δimpedance was a significant predictor of SP occurrence. “PV-side-of-ridge” ablation at antero-superior and roof segments of left PV might predispose to SP occurrence due to excessive tissue coverage.
No takes yet. Share an insight, caveat, or question.
Shuang et al. (2022) conducted a case-control in Pulmonary vein isolation (n=701). High-power circumferential pulmonary vein isolation was evaluated on Prevalence of steam pops. High-power circumferential pulmonary vein isolation resulted in steam pops in 1.57% of patients, with an impedance change >12 Ω serving as a significant predictor of occurrence.