Key result
Non-exercise VO2MAX estimation using the Seismofit sensor strongly correlates with CPET and yields similar mean values.
Why the study?
Cardiopulmonary exercise testing to determine maximal oxygen uptake is cumbersome and time-consuming, making it impractical in many testing scenarios.
Does a seismocardiography sensor accurately estimate maximal oxygen uptake compared to cardiopulmonary exercise testing in healthy subjects?
Cohort (n=94)
Does a seismocardiography sensor accurately estimate maximal oxygen uptake compared to cardiopulmonary exercise testing in healthy subjects?
Answers are grounded in this paper and cite their sources.
A novel seismocardiography sensor provides accurate and highly repeatable non-exercise estimates of VO2MAX that correlate well with standard cardiopulmonary exercise testing.
Background: The value of maximal oxygen uptake (VO2MAX) is a key health indicator. Usually, VO2MAX is determined with cardiopulmonary exercise testing (CPET), which is cumbersome and time-consuming, making it impractical in many testing scenarios. The aim of this study is to validate a novel seismocardiography sensor (Seismofit®, VentriJect DK, Hellerup, Denmark) for non-exercise estimation of VO2MAX. Methods: A cohort of 94 healthy subjects (52% females, 48.2 (8.7) years old) were included in this study. All subjects performed an ergometer CPET. Seismofit® measurements were obtained 10 and 5 min before CPET in resting condition and 5 min after exhaustion. Results: The CPET VO2MAX was 37.2 (8.6) mL/min/kg, which was not different from the two first Seismofit® estimates at 37.5 (8.1) mL/min/kg (p = 0.28) and 37.3 (7.8) mL/min/kg (p = 0.66). Post-exercise Seismofit® was 33.8 (7.1) mL/min/kg (p < 0.001). The correlation between the CPET and the Seismofit® was r = 0.834 and r = 0.832 for the two first estimates, and the mean average percentage error was 11.4% and 11.2%. Intraclass correlation coefficients between the first and second Seismofit® measurement was 0.993, indicating excellent test-retest reliability. Conclusion: The novel Seismofit® VO2MAX estimate correlates well with CPET VO2MAX, and the accuracy is acceptable for general health assessment. The repeatability of Seismofit® estimates obtained at rest was very high.
No takes yet. Share an insight, caveat, or question.
Schulenburg et al. (2024) conducted a cohort in Healthy subjects (n=94). Seismofit (Seismocardiography) vs. Cardiopulmonary exercise testing (CPET) was evaluated on Maximal oxygen uptake (VO2MAX) (r = 0.834, p=0.28). Non-exercise estimation of VO2MAX using the Seismofit sensor correlated strongly with cardiopulmonary exercise testing and yielded similar mean values.
Synapse has enriched 10 related papers. Consider them for comparative context: