Key result
Tissue harmonic imaging significantly increased the contrast-to-speckle ratio compared to fundamental imaging when an ethanol layer simulating human fat was present (p<0.05).
Why the study?
Does harmonic imaging improve cyst detectability compared to fundamental imaging in a tissue-mimicking phantom with and without a simulated fat layer?
Does harmonic imaging improve cyst detectability compared to fundamental imaging in a tissue-mimicking phantom with and without a simulated fat layer?
Answers are grounded in this paper and cite their sources.
Tissue harmonic imaging significantly improves image quality compared to fundamental imaging only in the presence of a simulated fat layer, suggesting its primary benefit is in obese rather than slim patients.
Tissue harmonic scanning visually improves echocardiographic image quality. The aim of the present study was to objectively assess the improvement in harmonic image quality under controlled laboratory conditions. A tissue-mimicking phantom that contained 8-mm-diameter cystic lesions at depths ranging from 2 to 12 cm was used. Harmonic scans (1.7 MHz transmit, 3.4 MHz receive) of the phantom were obtained and lesion detectability was compared to that in scans acquired with 2 fundamental frequencies (2.0 and 3.3 MHz). A 2 cm-thick ethanol layer was also used to simulate the nonlinear effect of human fat. Cyst detectability was quantified by measurement of the contrast-to-speckle ratio (CSR). The results indicated no significant difference in the CSR between harmonic and fundamental images obtained without the ethanol layer. With images obtained with the ethanol layer, a relative increase of the CSR during harmonic imaging was observed with respect to fundamental imaging (p<0.05). In conclusion, a fat layer, here simulated by ethanol, plays a significant role in determining the resulting image quality. Without this layer, the contribution of the second harmonic mode was not significant. Thus, in a slim patient, the harmonic mode may not be as beneficial to image improvement as in an obese patient.
No takes yet. Share an insight, caveat, or question.
Tanabe et al. (2000) studied this question. Harmonic scans (1.7 MHz transmit, 3.4 MHz receive) vs. Fundamental frequencies (2.0 and 3.3 MHz) was evaluated on Cyst detectability quantified by contrast-to-speckle ratio (CSR) (p=<0.05). Tissue harmonic imaging significantly increased the contrast-to-speckle ratio compared to fundamental imaging when an ethanol layer simulating human fat was present (p<0.05).