Key result
Spatial dispersion of Activation Recovery Interval extracted from unipolar recordings is consistent with optically recorded Action Potential Duration, validating ARI as a surrogate for APD.
Why the study?
The ability of the activation recovery interval to accurately predict action potential duration dispersion in patients with repolarization abnormality had not been determined.
Does Activation Recovery Interval (ARI) accurately reflect Action Potential Duration (APD) dispersion in human ventricles?
Observational (n=7)
No
Does Activation Recovery Interval (ARI) accurately reflect Action Potential Duration (APD) dispersion in human ventricles?
Activation Recovery Interval (ARI) accurately reflects underlying Action Potential Duration (APD) dispersion, validating its clinical use for assessing repolarization abnormalities.
Supports ARI as APD surrogate in human ventricles; leaves open larger trials before clinical integration.
Background: A large Dispersion of Repolarization (DoR) is associated with an increased arrhythmogenic risk.This can be measured clinically by calculating the Activation Recovery Interval (ARI) to estimate Action Potential Duration (APD).However, the ability of ARI to accurately predict APD dispersion in patients with repolarization abnormality has not been determined.Objective: Compare ARI calculated from patients with optical mapping of human hearts to establish the validity of ARI as a surrogate for APD.Methods: Optical mapping (OM) was performed on the left ventricles of 4 explanted human hearts.APD and repolarization times were measured endo-and epicardially on the anterior of the LV.Electroanatomic mapping was performed with CARTO over the entire endo-and epicardial surfaces of 3 patients.Activation and repolarization were calculated, dispersion of ARI was measured.Results: APD and ARI were consistent between mapping methods over most of the sub-regions studied.Epicardium ARI dispersion was consistently higher than that of the endocardium in both OM and CARTO datasets.Conclusion: APD distribution, and consequently DoR, agree between mapping methods.Measuring DoR by ARI accurately assesses the underlying repolarization abnormalities in patients.
No takes yet. Share an insight, caveat, or question.
Langfield et al. (2018) conducted an observational in Ventricular repolarization abnormalities (n=7). Activation Recovery Interval (ARI) calculation vs. Action Potential Duration (APD) from optical mapping was evaluated on Spatial dispersion of repolarization. Spatial dispersion of Activation Recovery Interval extracted from unipolar recordings is consistent with optically recorded Action Potential Duration, validating ARI as a surrogate for APD.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: