Key result
Minuteful Kidney Test detects albuminuria with ~96% sensitivity compared to standard lab methods in at-risk patients.
Why the study?
The diagnostic accuracy of the home-based semi-quantitative Minuteful Kidney Test compared to a standard laboratory method required evaluation for detecting albuminuria in people with or at risk for CKD.
Does the Minuteful Kidney Test accurately detect albuminuria compared to standard laboratory quantitative methods in individuals at risk for chronic kidney disease?
Cross-Sectional (n=555)
Blinded
Yes
Does the Minuteful Kidney Test accurately detect albuminuria compared to standard laboratory quantitative methods in individuals at risk for chronic kidney disease?
The smartphone-enabled Minuteful Kidney Test demonstrates high sensitivity and negative predictive value for detecting albuminuria, offering a reliable home-based screening tool for patients at risk of chronic kidney disease.
Answers are grounded in this paper and cite their sources.
May aid home albuminuria screening in at-risk groups; leaves open prospective validation before adoption.
Objective To evaluate the diagnostic accuracy of the FDA-cleared Minuteful Kidney Test (MKT), a home-based semi-quantitative urine albumin-to-creatinine ratio (uACR) test, compared to the Beckman Coulter AU 480 analyzer, a standard laboratory quantitative method, for detecting albuminuria in individuals with or at risk for chronic kidney disease (CKD). Methods Urine samples were obtained from individuals with risk factors for kidney damage. Each sample was analyzed for uACR using both the MKT and the laboratory standard method. Albuminuria was classified according to KDIGO guidelines: A1 (uACR ≤30 mg/g, “Normal to mildly increased”), A2 (uACR 30-300 mg/g, “Moderately increased”), and A3 (uACR ≥300 mg/g, “Severely increased”). Sensitivity, specificity, PPV, and NPV of MKT were calculated against the laboratory standard. Results Out of 615 collected urine samples, 555 met inclusion criteria. Albuminuria (uACR ≥30 mg/g) was detected in 24.9% of samples using the laboratory method (95% CI, 21.3–28.5%). The MKT demonstrated a sensitivity of 96.4%, specificity of 84.2%, NPV of 98.6%, and PPV of 66.8%. The test accurately categorized 85.6% of samples into the KDIGO albuminuria categories A1, A2, and A3, accurately identifying 100% of laboratory-confirmed A3 samples as abnormal. Conclusions The Minuteful Kidney Test shows high sensitivity, exceeding the guideline-required threshold (>85%, P<0.001), and robust specificity for detecting albuminuria in high-risk populations. This device offers potential to improve CKD screening and management and facilitate early intervention in primary care and community settings, where accurate rule-out capabilities are essential.
No takes yet. Share an insight, caveat, or question.
Jeddah et al. (2025) conducted a cross-sectional in At risk for chronic kidney disease (CKD) (n=555). Minuteful Kidney Test (MKT) vs. Beckman Coulter AU 480 analyzer was evaluated on Sensitivity for detecting albuminuria (uACR ≥30 mg/g) (Sensitivity 96.4%, 95% CI 93.3-99.5, p=<0.001). The Minuteful Kidney Test demonstrated a sensitivity of 96.4% and specificity of 84.2% for detecting albuminuria compared to a standard laboratory quantitative method in at-risk populations.