Key result
SPIHT algorithm achieves ~90-fold ECG compression while maintaining high AF detection accuracy.
Why the study?
ECG signal compression is employed to increase battery life and transmission efficiency in resource-constrained cardiac devices, but its impact on preserving the diagnostic accuracy of AF detection required investigation.
Does ECG signal compression using the SPIHT algorithm affect the accuracy of an entropy-based AF detection algorithm?
Does ECG signal compression using the SPIHT algorithm affect the accuracy of an entropy-based AF detection algorithm?
ECG signal compression up to a ratio of 90 preserves high diagnostic accuracy for AF detection, enabling significant energy savings for resource-constrained devices.
Although Atrial Fibrillation (AF) is the most frequent cause of cardioembolic stroke, the arrhythmia remains underdiagnosed, as it is often asymptomatic or intermittent. Automated detection of AF in ECG signals is important for patients with implantable cardiac devices, pacemakers or Holter systems. Such resource-constrained systems often operate by transmitting signals to a central server where diagnostic decisions are made. In this context, ECG signal compression is being increasingly investigated and employed to increase battery life, and hence the storage and transmission efficiency of these devices. At the same time, the diagnostic accuracy of AF detection must be preserved. This paper investigates the effects of ECG signal compression on an entropy-based AF detection algorithm that monitors R-R interval regularity. The compression and AF detection algorithms were applied to signals from the MIT-BIH AF database. The accuracy of AF detection on reconstructed signals is evaluated under varying degrees of compression using the state-of-the-art Set Partitioning In Hierarchical Trees (SPIHT) compression algorithm. Results demonstrate that compression ratios (CR) of up to 90 can be obtained while maintaining a detection accuracy, expressed in terms of the area under the receiver operating characteristic curve, of at least 0.9. This highlights the potential for significant energy savings on devices that transmit/store ECG signals for AF detection applications, while preserving the diagnostic integrity of the signals, and hence the detection performance.
No takes yet. Share an insight, caveat, or question.
Cervigón et al. (2021) studied Atrial Fibrillation. ECG signal compression using SPIHT algorithm vs. Varying degrees of compression was evaluated on Accuracy of AF detection (area under the receiver operating characteristic curve). ECG signal compression using the SPIHT algorithm achieved compression ratios up to 90 while maintaining an AF detection accuracy (AUC) of at least 0.9.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: