This is the authors' abstract. We don't add key points for this paper.
Radiation dose from mammographic techniques was determined as a function of surface exposure, beam quality, and depth. Relative exposure vs. depth was measured in tissue-substitute materials by thermoluminescent dosimetry. The f-factors were calculated from elemental compositions of mastectomy specimens. Dose at depth depends on beam quality as well as exposure and tissue composition. Analysis of data from the ACS/NCI Screening Centers shows current average midbreast doses to be 25 times lower (film/screen) and 3 times lower (Xerox) than the 2 rads previously estimated. Quantitative risk indicators other than midbreast dose are also discussed.