Key result
The CardiWall system achieved a high true positive rate of 91.4% and a very low false positive rate of 1% for detecting malicious clinical programming of cardiac implantable devices.
Why the study?
CIED programmer devices are open to external connections and vulnerable to cyber-attacks that could manipulate device operations and harm patients.
Can the CardiWall system accurately detect malicious clinical programming commands directed at cardiac implantable electronic devices?
Can the CardiWall system accurately detect malicious clinical programming commands directed at cardiac implantable electronic devices?
Answers are grounded in this paper and cite their sources.
The CardiWall system demonstrates high accuracy in detecting malicious programming commands for cardiac implantable electronic devices, offering a potential cybersecurity safeguard.
Today, cardiac implantable electronic devices (CIEDs), such as pacemakers and implantable cardioverter defibrillators (ICDs), play an increasingly important role in healthcare ecosystems as patient life support devices. Physicians control, program and configure CIEDs on a regular basis using a dedicated programmer device. The programmer device is open to external connections (e.g., USB, Bluetooth, etc.), and thus it is exposed to a variety of cyber-attacks by which an attacker can manipulate the programmer device's operations and consequently harm the patient. In this paper, we present CardiWall, a novel detection and prevention system designed to protect ICDs from cyber-attacks aimed at the programmer device. Our system has six different layers of protection, leveraging medical experts' knowledge, statistical methods, and machine learning algorithms. We evaluated the CardiWall system extensively in two comprehensive experiments. For the evaluation, we gathered data for a period of four years and used 775 benign clinical commands that are related to hundreds of different patients (obtained from different programmer devices located at Barzilai University Medical center) and 28 malicious clinical commands (created by two cardiology experts from different hospitals). The evaluation results show that only two out of the six layers proposed in CardiWall system provided a high detection capability associated with high rates of true positive, and low rates of false positive. With the configuration that provided the best harmonic mean of sensitivity and specificity (HMSS), CardiWall achieved a high true positive rate (TPR) of 91.4% and a very low false positive rate (FPR) of 1%, with an AUC of 94.7%.
No takes yet. Share an insight, caveat, or question.
Kintzlinger et al. (2020) studied Cardiac implantable electronic devices (CIEDs) cyber-attacks. CardiWall system was evaluated on Detection capability (True positive rate, false positive rate, AUC) (AUC 94.7%). The CardiWall system achieved a high true positive rate of 91.4% and a very low false positive rate of 1% for detecting malicious clinical programming of cardiac implantable devices.
Synapse has enriched 10 related papers. Consider them for comparative context: