Key result
Voltage mapping using a real-time MRI-guided electrophysiology system identified MR-derived late gadolinium enhancement with 57% sensitivity and 96% specificity in a porcine model.
Why the study?
Real-time MRI-guided electrophysiology offers potential advantages for substrate and ablation assessment, but a novel system needed evaluation for endocardial voltage and delayed conduction mapping.
Does a real-time MR-EP system accurately identify structural and electrophysiological ventricular tachycardia substrate compared to LGE-MRI in a porcine model?
Does a real-time MR-EP system accurately identify structural and electrophysiological ventricular tachycardia substrate compared to LGE-MRI in a porcine model?
A novel real-time MRI-guided electrophysiology system demonstrated moderate accuracy in identifying ventricular scar and delayed conduction compared to LGE-MRI in a porcine model.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for integrated MR-EP systems; leaves open human validation for VT substrate mapping.
AIMS: Potential advantages of real-time magnetic resonance imaging (MRI)-guided electrophysiology (MR-EP) include contemporaneous three-dimensional substrate assessment at the time of intervention, improved procedural guidance, and ablation lesion assessment. We evaluated a novel real-time MR-EP system to perform endocardial voltage mapping and assessment of delayed conduction in a porcine ischaemia-reperfusion model. METHODS AND RESULTS: Sites of low voltage and slow conduction identified using the system were registered and compared to regions of late gadolinium enhancement (LGE) on MRI. The Sorensen-Dice similarity coefficient (DSC) between LGE scar maps and voltage maps was computed on a nodal basis. A total of 445 electrograms were recorded in sinus rhythm (range: 30-186) using the MR-EP system including 138 electrograms from LGE regions. Pacing captured at 103 sites; 47 (45.6%) sites had a stimulus-to-QRS (S-QRS) delay of ≥40 ms. Using conventional (0.5-1.5 mV) bipolar voltage thresholds, the sensitivity and specificity of voltage mapping using the MR-EP system to identify MR-derived LGE was 57% and 96%, respectively. Voltage mapping had a better predictive ability in detecting LGE compared to S-QRS measurements using this system (area under curve: 0.907 vs. 0.840). Using an electrical threshold of 1.5 mV to define abnormal myocardium, the total DSC, scar DSC, and normal myocardium DSC between voltage maps and LGE scar maps was 79.0 ± 6.0%, 35.0 ± 10.1%, and 90.4 ± 8.6%, respectively. CONCLUSION: Low-voltage zones and regions of delayed conduction determined using a real-time MR-EP system are moderately associated with LGE areas identified on MRI.
No takes yet. Share an insight, caveat, or question.
Mukherjee et al. (2019) studied Ventricular tachycardia substrate (myocardial infarction) (n=7). Real-time MRI-guided electrophysiology (MR-EP) system vs. Late gadolinium enhancement (LGE) MRI was evaluated on Sensitivity of voltage mapping to identify MR-derived LGE using conventional (0.5mV-1.5mV) bipolar voltage thresholds. Voltage mapping using a real-time MRI-guided electrophysiology system identified MR-derived late gadolinium enhancement with 57% sensitivity and 96% specificity in a porcine model.
Synapse has enriched 10 related papers. Consider them for comparative context: