Demonstrates improved sensitivity for hydrogen peroxide detection in various applications, indicating its environmental relevance.
Nonenzymatic electrochemical methods offer significant advantages for hydrogen peroxide (H2O2) detection. However, the sensitivity must be improved for practical applications. In this study, we developed a novel electrochemical sensor for H2O2 based on a Cu-TCPP/reduced graphene oxide (rGO) nanocomposite synthesized via a surfactant-assisted method. Comprehensive structural and morphological analyses, including TEM, SEM, EDS, XRD, UV–vis, and FTIR, demonstrated uniform dispersion of Cu-TCPP on the rGO surface via π–π interactions. Electrocatalytic reduction of H2O2 at the Cu-TCPP/rGO-modified electrode showed a linear range of 0.10–6.30 mM and a detection limit of 0.03 mM (S/N = 3). The practical applicability of the sensor was demonstrated by the detection of H2O2 in commercial disinfectant samples, yielding recovery rates in the range of 99.7–104.0%. These results highlight the capabilities of Cu-TCPP/rGO nanocomposites as reliable platforms for the sensitive, nonenzymatic electrochemical detection of H2O2, making them a valuable potential tool for environmental monitoring.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: