Answers are grounded in this paper and cite their sources.
Observational study demonstrates rapid confocal microscopy detects tumour cells in neurosurgical patients, indicating strong potential for real-time surgical guidance.
Real-time intra-operative brain tumour tissue analysis can reduce turnaround times and enable repeated sampling, enhancing diagnostic accuracy and guiding resection. We investigated the use of an ultra-fast confocal microscopy scanner (Histolog®, SamanTree Medical SA) for real-time brain tumour tissue diagnosis. This study aims to demonstrate Histolog®’s diagnostic accuracy in a cohort of brain or spinal cord tumour patients. A single-group, observational study included adult patients undergoing tissue biopsies or tumour debulking surgery. Multiple, freshly excised tissue samples were stained and imaged within 60 seconds using Histolog® alongside standard diagnostic methods. A Consultant Neuropathologist performed a blinded concordance analysis. Of 47 specimens, 94% produced interpretable images during surgery. Histolog® images enabled real-time tumour cell identification in all specimens and matched gold standard diagnoses in 68% of cases. Histolog® demonstrated rapid intraoperative imaging with diagnostic performance for tumour presence that was comparable to existing intraoperative methods in this small cohort, while exact histological concordance was lower (68%). Future studies will explore real-time margin zone analysis and machine learning for automatic diagnosis.
No takes yet. Share an insight, caveat, or question.
Bolton et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: