A programmable ECG generator based on a dynamical model implemented on a microcontroller is feasible for testing, calibrating, and maintaining electrocardiograph equipment.
Answers are grounded in this paper and cite their sources.
Supports ECG equipment calibration and maintenance; extends McSharry model but leaves clinical validation open.
This article reports on the design and development of a complete, programmable electrocardiogram (ECG) generator, which can be used for the testing, calibration and maintenance of electrocardiograph equipment. A modified mathematical model, developed from the three coupled ordinary differential equations of McSharry et al. [IEEE Trans. Biomed. Eng. 50, 289, (2003)], was used to locate precisely the positions of the onset, termination, angle, and duration of individual components in an ECG. Generator facilities are provided so the user can adjust the signal amplitude, heart rate, QRS-complex slopes, and P- and T-wave settings. The heart rate can be adjusted in increments of 1BPM (beats per minute), from 20to176BPM, while the amplitude of the ECG signal can be set from 0.1to400mV with a 0.1mV resolution. Experimental results show that the proposed concept and the resulting system are feasible.