C beam. The high resolution dosimetric measurements show a good agreement between simulated and measured dose distributions. Irradiation of the monoenergetic raster plan took 3 s, which is approximately 20 times shorter than a comparable plan with 16 different energies. The combination of only one energy and a 3D range-modulator leads to a tremendous decrease in irradiation time. 'Interplay effects', typical for moving targets and pencil beam scanning, can be immensely reduced or disappear completely, making the delivery of a homogeneous dose to moving targets more reliable. Combining high dose conformity, very good homogeneity and extremely short irradiation times, the 3D range-modulator is considered to become a clinically applicable method for very fast treatment of lung tumours.