Why the study?
Validating AF source mapping systems is confounded by the absence of ground truth, and a key concern of prior mapping technologies is poor spatiotemporal map reproducibility.
Does Electrographic flow (EGF) mapping produce reproducible intra- and inter-procedural maps in patients with persistent atrial fibrillation?
Does Electrographic flow (EGF) mapping produce reproducible intra- and inter-procedural maps in patients with persistent atrial fibrillation?
Electrographic flow mapping demonstrates high intra- and inter-procedural reproducibility for identifying atrial fibrillation sources, providing a stable target for ablation.
Answers are grounded in this paper and cite their sources.
Reproducible electrographic flow maps in persistent AF support further evaluation; leaves open clinical utility pending randomized trials.
AIMS: Validating mapping systems that identify atrial fibrillation (AF) sources (focal/rotational activity) is confounded by the absence of ground truth. A key concern of prior mapping technologies is spatiotemporal instability, manifesting as poor map reproducibility. Electrographic flow (EGF) employs a novel algorithm that visualizes atrial electrical wavefront propagation to identify putative AF sources. We analysed both intra- (3 min) and inter- (>3 months) procedure EGF map reproducibility. METHODS AND RESULTS: In 23 persistent AF patients, after pulmonary vein isolation (PVI), EGF maps were generated from 3 serial 1 min recordings using a 64-electrode basket mapping catheter (triplets) at right and left atrial locations. Source prevalence from map triplets was compared between recordings. Per protocol, 12 patients returned for 3-month remapping (1 non-inducible): index procedure post-PVI EGF maps were compared with initial EGF remapping at 3-month redo. Intra-procedure reproducibility: analysing 224 map triplets (111 right atrium, 113 left atrium) revealed a high degree of map consistency with minimal min-to-min shifts: 97 triplets (43%), exact match of leading sources on all 3 maps; 95 triplets (42%), leading source within 1 electrode space on 2 of 3 maps; and 32 triplets (14%), chaotic leading source pattern. Average deviation in source prevalence over 60 s was low (6.4%). Inter-procedure reproducibility: spatiotemporal stability of EGF mapping >3 months was seen in 16 of 18 (89%) sources mapped in 12 patients with (re)inducible AF. CONCLUSION: Electrographic flow mapping generates reproducible intra- and inter-procedural maps, providing rationale for randomized clinical trials targeting these putative AF sources.
No takes yet. Share an insight, caveat, or question.
Reddy et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: