Why the study?
The risks of damage to the S-ICD electrode during sternotomy and adverse interactions with sternal wires remained unclear.
Does median sternotomy before or after S-ICD implantation increase the risk of lead damage, inappropriate shocks, or failure to terminate arrhythmias compared to nonsternotomy patients?
Does median sternotomy before or after S-ICD implantation increase the risk of lead damage, inappropriate shocks, or failure to terminate arrhythmias compared to nonsternotomy patients?
Median sternotomy performed before or after S-ICD implantation is safe and effective, with no increased risk of lead damage, inappropriate shocks, or device failure compared to patients without sternotomy.
Supports sternotomy compatibility with S-ICD without excess risk; extends observational data but leaves open need for prospective confirmation.
BACKGROUND: Subcutaneous implantable cardioverter defibrillators (S-ICDs) are an attractive alternative to transvenous ICDs among those not requiring pacing. However, the risks of damage to the S-ICD electrode during sternotomy and adverse interactions with sternal wires remain unclear. We sought to determine the rates of damage to the S-ICD lead during sternotomy, inappropriate shocks from electrical noise due to interaction with sternal wires, and failure to terminate spontaneous or induced ventricular arrhythmias. METHODS: Retrospective, multicenter study of patients undergoing sternotomy before or after S-ICD implantation. Clinical, procedural, and device-related data were collected by each center and analyzed by the coordinating center. These data were compared with a historical control cohort of nonsternotomy patients. RESULTS: Of 196 identified patients (52±16 years, 47 women), 166 underwent S-ICD implantation after sternotomy and 30 sternotomy after S-ICD. There was no damage to any lead among those who underwent sternotomy after S-ICD. Defibrillation threshold testing was performed in 63% at implant, with 91% first shock success. During a median follow-up of 29 months (range, 1-188), S-ICD first shocks successfully terminated spontaneous ventricular arrhythmias in 31 of 32 patients (97%). Inappropriate shocks occurred in 22 patients, most commonly related to T wave oversensing (n=14). Compared with the nonsternotomy controls, there were no differences in rates of first shock success for induced or spontaneous arrhythmias or rate of inappropriate shocks. CONCLUSIONS: Sternotomy before or after S-ICD does not confer additional risk relative to a historical control group without sternotomy.
No takes yet. Share an insight, caveat, or question.
Sugrue et al. (2023) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: