Answers are grounded in this paper and cite their sources.
This study developed a new approach to produce sterile, hyperpolarized [ 13 C, 15 N 2 ]urea as a novel molecular imaging probe and applied it for first-ever healthy brain volunteer studies. Hyperpolarized [ 13 C, 15 N 2 ]urea, as a small, metabolically inert molecule, offers significant advantages for perfusion imaging due to its endogenous nature and excellent safety profile. The developed methods achieved a hyperpolarized [ 13 C, 15 N 2 ]urea solution (132 ± 6 mM) with 27.4 ± 5.6% polarization and a T 1 = 50.4 ± 0.2 s. In healthy brain volunteer studies, high-resolution 13 C imaging captured blood flow with a spatial resolution of 7.76 × 7.76 × 15 (or 10) mm 3 over ~1 min following hyperpolarized [ 13 C, 15 N 2 ]urea injection, visualizing detailed vascular structures. Time-to-peak and centroid analyses showed consistent arterial and venous signal patterns across subjects. Findings suggest hyperpolarized [ 13 C, 15 N 2 ]urea may have applications beyond brain imaging, including the non-invasive perfusion assessment in various organs, cancer microenvironment, and renal function, paving the way for clinical translation.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2025) studied this question.