Key result
Repositioning defibrillation catheters to the SVC achieves ~93% cardioversion success after failed CS positioning.
Why the study?
Achieving standard CS/RA intracardiac defibrillation catheter positioning via the IVC during AF ablation can be challenging, and alternative configurations remain underexplored.
Does repositioning an intracardiac defibrillation catheter to the SVC configuration improve cardioversion success compared to a CS-only configuration in patients undergoing AF ablation?
Observational (n=81)
Does repositioning an intracardiac defibrillation catheter to the SVC configuration improve cardioversion success compared to a CS-only configuration in patients undergoing AF ablation?
Repositioning an intracardiac defibrillation catheter to the SVC configuration after a failed CS-only attempt achieves a high rate of successful cardioversion without complications during AF ablation.
Answers are grounded in this paper and cite their sources.
May support SVC repositioning after CS failure in AF ablation; leaves open randomized confirmation.
Background In pulmonary vein isolation (PVI) for atrial fibrillation (AF), intraoperative defibrillation is often required. Intracardiac defibrillation catheters (ICDCs) are most effective when positioned to enclose the heart between the coronary sinus (CS) and right atrium (RA) (CS/RA configuration). However, achieving this positioning via the inferior vena cava (IVC) can be challenging, and alternative configurations remain underexplored. Methods This study included patients with paroxysmal or persistent AF who underwent cryoballoon ablation followed by intracardiac cardioversion using an ICDC via the IVC. The catheter was initially positioned with distal electrodes in the CS and proximal electrodes in the IVC (CS‐only configuration). If cardioversion failed, the catheter was repositioned to place distal electrodes in the superior vena cava (SVC configuration). A maximum of 30 J of energy was used for all cardioversion attempts. Results A total of 81 patients were included. Cardioversion in the CS‐only configuration restored sinus rhythm in 11% (9/81) of patients. Repositioning to the SVC configuration achieved successful cardioversion in 93.1% (67/72) of the remaining cases without complications. Patients requiring the SVC configuration had a significantly higher prevalence of persistent AF (33.3% vs. 80.6%; p = 0.045). No adverse events were observed following cardioversion in the SVC configuration. Conclusions While the CS‐only configuration offers ease of placement, its efficacy is limited. Repositioning to the SVC configuration significantly enhances cardioversion success and represents a safer, more effective alternative for ICDC use during AF ablation.
No takes yet. Share an insight, caveat, or question.
Ohashi et al. (2025) conducted an observational in Atrial fibrillation (n=81). Intracardiac defibrillation catheter in SVC configuration vs. CS-only configuration was evaluated on Successful cardioversion to sinus rhythm. Repositioning intracardiac defibrillation catheters to a superior vena cava configuration achieved successful cardioversion in 93.1% of patients who failed initial coronary sinus-only positioning.