January 20, 2009Journal of Hypertension210 citations

Use of automated office blood pressure measurement to reduce the white coat response

View Full Paper
MMMartin G. MyersGeneral CardiologyMVMiguel ValdiviesoClinica Universidad de NavarraAKAlexander KissSunnybrook Health Science Centre

Key Points

  • To evaluate whether using an automated sphygmomanometer while patients rest alone in an exam room reduces the white coat response compared to routine physician manual readings.
  • Studied 309 consecutive patients referred for hypertension evaluation at an academic ambulatory blood pressure (ABP) monitoring unit.

Structured PICO

Does automated office blood pressure measurement using the BpTRU device reduce the white coat response compared to manual physician measurement in patients referred for hypertension?

P
Population
309 consecutive patients referred for diagnosis or management of hypertension to an ambulatory blood pressure monitoring unit in an academic hospital.
I
Intervention
Automated office blood pressure measurement using the BpTRU device with the patient resting alone in a quiet examining room.
C
Comparator
Blood pressure readings taken by the patient's own family physician using a manual sphygmomanometer.
O
Outcome
Differences between mean awake ambulatory blood pressure (ABP) and office BP readings (automated vs manual).surrogate

Automated office blood pressure measurement with the patient resting alone significantly reduces the white coat response compared to routine manual measurement by a physician.

Abstract

OBJECTIVE: To examine the possibility of reducing the white coat response using an automated sphygmomanometer designed for office use, the BpTRU. Consecutive patients referred from physicians in the community to an ambulatory blood pressure (ABP) monitoring unit in an academic hospital were included in the study. PARTICIPANTS AND METHODS: A total of 309 patients referred for diagnosis or management of hypertension were studied. Differences between mean awake ABP and BP readings taken by the patient's own physician using a manual sphygmomanometer or the automated BpTRU device with the patient resting alone in the ABP monitoring unit were compared. RESULTS: BP recorded in the examining room using an automated device (132 +/- 19/75 +/- 12) was similar to the mean awake ABP (134 +/- 12/77 +/- 10) with both values being lower (P < 0.001) than the BP recorded on a routine visit to the patient's own family physician (152 +/- 18/87 +/- 11). The coefficient of correlation between the systolic/diastolic ABP and the automated office BP (r = 0.62/0.72) was higher (P < 0.001) than with the family physician's manual BP (r = 0.32/0.48). The prevalence of white coat hypertension in untreated patients (n = 146) was significantly (P < 0.001) lower with automated office BP (16%) compared with the routine family physician BP (55%). CONCLUSION: The white coat response associated with office BP measurements can be virtually eliminated by recording BP with the automated BpTRU device with patients resting alone in a quiet examining room.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Myers et al. (2009) studied this question.

synapsesocial.com/papers/6a19d9563e4c9aaeb7f670cchttps://doi.org/10.1097/hjh.0b013e32831b9e6b
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1White Coat Phenomenon in Patients Receiving Antihypertensive Therapy1991 · 86 citations
  2. 2Evaluation of the overall efficacy of the Omron office digital blood pressure HEM-907 monitor in adults2001 · 217 citations
  3. 3Prevalence, Causes, and Consequences of Masked Hypertension: A Meta-analysis2008 · 145 citations
  4. 4Comparison of two automated sphygmomanometers for use in the office setting2009 · 42 citations
  5. 5Optimum frequency of office blood pressure measurement using an automated sphygmomanometer2008 · 57 citations