Answers are grounded in this paper and cite their sources.
A unique method using membrane preconcentration capillary electrophoresis (mPC-CE) on-line with tandem mass spectrometry (mPC-CE-MS/MS) for sequencing <100 fmols of a series of MHC class I synthetic peptides is described. We show how this methodology in conjunction with transient isotachophoresis (tlTP) after analyte elution from the membrane improves conventional CE-MS strategies by permitting the analysis of large sample volumes (>100 μL) while maintaining high analyte resolution and separation efficiencies that are typically afforded by CE. Instrumental parameters, including the internal dimensions of the mPC-CE capillary, collision gas pressure, collision gas type, and collision energy, are all shown to have substantial effect on the abundance of product ions produced in tandem MS/MS spectra when attempting to sequence peptides at the sub 100 femtomole level. Furthermore, we demonstrate that the physico-chemical properties of these analytes can also affect the MS/MS product ion abundance. In particular, the presence of acidic residue tends to reduce mPC-tlTP-CE-MS/MS sensitivity and direct fragmentation processes resulting in incomplete sequence information. This can be overcome by a simple esterification of the carboxylic acid functional group. Ultimately, we present an optimized mPC-tlTP-CE-MS/MS approach that permits peptide sequence determination by consumption of ∼40 femtomole of peptide analyte.
No takes yet. Share an insight, caveat, or question.
Tomlinson et al. (1995) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: