Key result
Rigid IVC filter models misestimate key hemodynamic parameters by up to ~12% versus hyperelastic models.
Why the study?
The accuracy of two common simplifying assumptions in numerical studies of IVC filters—the rigid wall assumption and the laminar flow model—needed examination compared with more realistic models.
In computational simulations of IVC filters, the rigid wall assumption significantly alters hemodynamic predictions compared to a hyperelastic model, particularly in the presence of large clots.
Rigid-wall models may misestimate IVC filter hemodynamics with clots; leaves open whether hyperelastic modeling improves simulation accuracy for device research.
Inferior vena cava (IVC) filters are vital in preventing pulmonary embolism (PE) by trapping large blood clots, especially in patients unsuitable for anticoagulation. In this study, the accuracy of two common simplifying assumptions in numerical studies of IVC filters-the rigid wall assumption and the laminar flow model-is examined, contrasting them with more realistic hyperelastic wall and turbulent flow models. Using fluid-structure interaction (FSI) and computational fluid dynamics (CFD) techniques, the investigation focuses on three hemodynamic parameters: time-averaged wall shear stress (TAWSS), oscillatory shear index (OSI), and relative residence time (RRT). Simulations are conducted with varying sizes of clots captured in the filter. The findings show that, in regions of high wall shear stress, the rigid wall model predicted higher TAWSS values, suggesting an increased disease risk compared to the hyperelastic model. However, the laminar and turbulent flow models did not show significant differences in TAWSS predictions. Conversely, in areas of low wall shear stress, the rigid wall model indicated lower OSI and RRT, hinting at a reduced risk compared to the hyperelastic model, with this discrepancy being more evident with larger clots. While the predictions for OSI and TAWSS were closely aligned for both laminar and turbulent flows, divergences in RRT predictions became apparent, especially in scenarios with very large clots.
No takes yet. Share an insight, caveat, or question.
Moradicheghamahi et al. (2024) studied Pulmonary embolism (Inferior vena cava filter hemodynamics). Hyperelastic wall and turbulent flow models vs. Rigid wall and laminar flow models was evaluated on Hemodynamic parameters including time-averaged wall shear stress (TAWSS), oscillatory shear index (OSI), and relative residence time (RRT). In computational simulations of IVC filters with large clots, the rigid wall model deviated from the hyperelastic model by up to 6.05% for mean TAWSS, 4.4% for mean OSI, and 12.06% for mean RRT.