Key result
A prototype series of active electrodes manufactured in thick-film technology demonstrated high immunity to interference, resulting in very high quality ECG and EEG recordings.
A newly developed low-cost active electrode using thick-film technology improves the resolution and interference immunity of biopotential recordings such as ECG.
Answers are grounded in this paper and cite their sources.
Thick-film active electrodes merit further validation; extends interference models but leaves open clinical adoption.
The accepted models of the origin of interference in body surface potential recordings predict that it would be advantageous to perform signal amplification on the electrode. A suitable circuit for a miniature biopotential amplifier to be manufactured in thick-film technology is described. A prototype series of active electrodes was produced and tested. The high immunity to interference resulted in ECG and EEG recordings of a very high quality.
No takes yet. Share an insight, caveat, or question.
MettingVanRijn et al. (2002) studied Biopotential recordings (ECG and EEG). Low-cost active electrode was evaluated on Immunity to interference and recording quality. A prototype series of active electrodes manufactured in thick-film technology demonstrated high immunity to interference, resulting in very high quality ECG and EEG recordings.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: