Key result
Volumetric MCE closely tracks PET myocardial blood flow with an r2 of ~0.92.
Why the study?
Does myocardial contrast echocardiography using a volumetric model accurately quantify absolute myocardial blood flow compared to PET and intracoronary Doppler in humans?
Population
20 healthy volunteers and 15 patients with coronary artery disease (coronary stenosis 0% to 89%). Total n=35.
Comparison
Myocardial contrast echocardiography using a… vs Positron emission tomography in healthy…
Design
Cross-sectional
Authors
Loading...
A volumetric model of ultrasound contrast agent kinetics allows for the noninvasive quantification of absolute myocardial blood flow using myocardial contrast echocardiography, showing strong correlation with PET and intracoronary Doppler.
Observational (n=35)
Does myocardial contrast echocardiography using a volumetric model accurately quantify absolute myocardial blood flow compared to PET and intracoronary Doppler in humans?
A volumetric model of ultrasound contrast agent kinetics allows for the noninvasive quantification of absolute myocardial blood flow using myocardial contrast echocardiography, showing strong correlation with PET and intracoronary Doppler.
Vogel et al. (2005) conducted an observational in Coronary artery disease and healthy volunteers (n=35). Myocardial contrast echocardiography (MCE) vs. Positron emission tomography (PET) and intracoronary Doppler was evaluated on Myocardial blood flow (MBF) quantification. Myocardial contrast echocardiography using a volumetric model showed good agreement with PET for resting and hyperemic perfusion (r2=0.92) and with Doppler for flow velocity reserve (r2=0.73).