Answers are grounded in this paper and cite their sources.
OBJECTIVE: The objective of the study was to determine accuracy of the Tanita SC-240 body composition analyser to measure paediatric percent body fat (%BF). METHODS: Eighty-nine African-American and white 5-18-year-olds participated in this study. %BF was estimated by dual-energy X-ray absorptiometry (DXA) and by the Tanita SC-240. RESULTS: Overall %BF was 33.5 ± 10.5% (Tanita SC-240) vs. 34.5 ± 8.7% (DXA). There was no significant difference between the two measures (P = 0.52, average error = -1.0%, average absolute error = 3.9%). The Tanita mean %BF estimates significantly differed from the DXA mean %BF in white boys (P = 0.001, Cohen's d = 0.40) and white girls (P = 0.006, Cohen's d = 0.48), but differences were of small effect. No differences in %BF estimates were found for African-American boys or girls. CONCLUSIONS: In this sample, the Tanita SC-240 demonstrated acceptable accuracy for estimating %BF when compared with DXA, supporting its use in field studies.
No takes yet. Share an insight, caveat, or question.
Barreira et al. (2012) studied this question.