Key result
KairoSight-3.0 software generated optical maps with high accuracy, with >97% of software and manual values falling within 10 ms of each other for action potential and calcium transient durations.
Why the study?
Analysis of complex optical datasets from cardiac optical mapping is time intensive and technically challenging.
Does KairoSight-3.0 software accurately measure cardiac electrophysiology parameters compared to manual mapping in isolated animal hearts?
Does KairoSight-3.0 software accurately measure cardiac electrophysiology parameters compared to manual mapping in isolated animal hearts?
KairoSight-3.0 is a validated, semi-automated software package that accurately characterizes cardiac electrophysiology and calcium handling from optical mapping data.
Answers are grounded in this paper and cite their sources.
Accelerates preclinical optical mapping analysis in animal hearts; leaves open translation to clinical electrophysiology.
Background Cardiac optical mapping is an imaging technique that measures fluorescent signals across a cardiac preparation. Dual optical mapping of voltage-sensitive and calcium-sensitive probes allow for simultaneous recordings of cardiac action potentials and intracellular calcium transients with high spatiotemporal resolution. The analysis of these complex optical datasets is both time intensive and technically challenging; as such, we have developed a software package for semi-automated image processing and analysis. Herein, we report an updated version of our software package ( KairoSight-3 . 0 ) with features to enhance characterization of cardiac parameters using optical signals. Methods To test software validity and applicability, we used Langendorff-perfused heart preparations to record transmembrane voltage and intracellular calcium signals from the epicardial surface. Isolated hearts from guinea pigs and rats were loaded with a potentiometric dye (RH237) and/or calcium indicator dye (Rhod-2AM) and fluorescent signals were acquired. We used Python 3.8.5 programming language to develop the KairoSight-3 . 0 software. Cardiac maps were validated with a user-specified manual mapping approach. Results Manual maps of action potential duration (30 or 80% repolarization), calcium transient duration (30 or 80% reuptake), action potential and calcium transient alternans were constituted to validate the accuracy of software-generated maps. Manual and software maps had high accuracy, with >97% of manual and software values falling within 10 ms of each other and >75% within 5 ms for action potential duration and calcium transient duration measurements (n=1000-2000 pixels). Further, our software package includes additional cardiac metric measurement tools to analyze signal-to-noise ratio, conduction velocity, action potential and calcium transient alternans, and action potential-calcium transient coupling time to produce physiologically meaningful optical maps. Conclusions KairoSight-3 . 0 has enhanced capabilities to perform measurements of cardiac electrophysiology, calcium handling, and the excitation-contraction coupling with satisfactory accuracy. Graphical Abstract Demonstrating Experimental and Data Analysis Workflow Created with Biorender.com
No takes yet. Share an insight, caveat, or question.
Haq et al. (2023) studied this question. KairoSight-3.0 software vs. Manual mapping approach was evaluated on Accuracy of software-generated maps compared to manual maps. KairoSight-3.0 software generated optical maps with high accuracy, with >97% of software and manual values falling within 10 ms of each other for action potential and calcium transient durations.