Answers are grounded in this paper and cite their sources.
A new hybrid linear ion trap time-of-flight mass spectrometer has been constructed and the first results for tandem-in-time MS/MS in a 2-dimensional Paul trap are presented. The instrument consists of an electrospray ionization source and two radio frequency (RF) only quadrupoles operating at a pressure of 7 mTorr N2 which are coupled orthogonally to a linear TOFMS. Ions are trapped in an RF-only quadrupole through the application of timed stopping potentials on the entrance and exit apertures. Through the coupling of an auxiliary waveform generator on one pair of the ion trap rods a small dipolar excitation voltage is superimposed on the quadrupolar electric field. This voltage is used for precursor isolation via broadband excitation and collision induced dissociation through mass selective resonant excitation. Instrumental performance and choice of operating parameters are demonstrated by recording fragment spectra for +3 ions of renin substrate and +1 ions of reserpine. TOFMS mass resolution (m/Δm)FWHM is as high as 740 at m/z = 609. Fragmentation efficiency is greater than 50% and excitation mass resolution is 60–75.
No takes yet. Share an insight, caveat, or question.
Campbell et al. (1998) studied this question.