Key result
64-slice CTA is linked to ~72% higher estimated radiation dose versus 16-slice CTA.
Why the study?
The increase in spatial and temporal resolution with 64-slice CTA is associated with increased radiation dose, and the impact of dose-saving protocols on radiation exposure and image quality in daily practice needed investigation.
Does 64-slice CTA compared with 16-slice CTA increase radiation dose, and do dose-saving algorithms reduce this exposure without compromising image quality in patients with suspected coronary artery disease?
Observational (n=1,035)
Does 64-slice CTA compared with 16-slice CTA increase radiation dose, and do dose-saving algorithms reduce this exposure without compromising image quality in patients with suspected coronary artery disease?
While 64-slice coronary CTA increases radiation exposure compared to 16-slice CTA, the use of ECG-dependent dose modulation and reduced tube voltage can effectively mitigate this risk without loss of diagnostic image quality.
64-slice CTA raises radiation exposure versus 16-slice; leaves open optimal dose-reduction protocols in routine coronary CTA.
BACKGROUND: Multislice computed tomography angiography (CTA) is a promising technology for imaging patients with suspected coronary artery disease. Compared with 16-slice CTA, the improved spatial and temporal resolution of 64-slice CTA (0.6- versus 1.0-mm slice thickness and 330- versus 420-ms gantry rotation time) is associated with an increase in radiation dose. The objective of this retrospective investigation was to compare the estimated dose received during 16- and 64-slice CTA in daily practice and to investigate the impact of different scan protocols on dose and image quality. METHODS AND RESULTS: Radiation dose was estimated for 1035 patients undergoing coronary CTA. Scanning algorithms with and without an ECG-dependent dose modulation and with a reduced tube voltage were investigated on dose estimates and image quality. In the entire patient cohort, radiation dose estimates were 6.4+/-1.9 and 11.0+/-4.1 mSv for 16- and 64-slice CTA, respectively (P<0.01). The reduction in radiation dose estimates ranged between 37% and 40% and between 53% and 64% with the use of ECG-dependent dose modulation and with the combined use of the dose modulation and a reduced tube voltage, respectively. The reduction in dose estimates was not associated with a reduction in diagnostic image quality as assessed by the signal-to-noise ratio and by the frequency of coronary segments with diagnostic image quality. CONCLUSIONS: The increase in spatial and temporal resolution with 64-slice CTA is associated with an increased radiation dose for coronary CTA. Dose-saving algorithms are very effective in reducing radiation exposure and should be used whenever possible.
No takes yet. Share an insight, caveat, or question.
Hausleiter et al. (2006) conducted an observational in suspected coronary artery disease (n=1,035). 64-slice CTA vs. 16-slice CTA was evaluated on Radiation dose estimates (p=<0.01). 64-slice CTA was associated with a significantly higher estimated radiation dose compared to 16-slice CTA (11.0 vs 6.4 mSv, P<0.01), though dose-saving algorithms effectively reduced exposure.