With an optical interferometer, the free vibration spectra and the local vibration amplitude of four rectangular atomic force microscope cantilevers made of silicon have been examined experimentally in a spectral range of 100 kHz to 10 MHz. A good agreement with the flexural wave theory of elastic beams was found. Coupling to torsional vibrations was also observed. When the sensor tip of the cantilever is in contact with a sample surface the resonances are shifted in frequency and the vibration amplitudes along the cantilever change. A method is presented to calculate this frequency shift using a linear approximation for the tip–sample interaction forces, and the results are compared with the frequency shift calculated from the point-mass model. The measured resonance frequencies of a surface-coupled cantilever do not correspond as well to the theoretical ones as in the free case even if the elastic-beam model is used. The reason for the disagreement is found to be the geometry of the commercial cantilevers and the nonlinearity of the tip-sample interaction force.